Like other devices, medical implants such as pacemakers are also getting smarter. For instance, in many cases these devices are connected to the cloud to transmit data from patients to doctors, and the devices can be programmed via special programming devices called 'pacemaker programmers'. This obviously offers all kinds of advantages, but unfortunately also comes with drawbacks. For instance, researchers from the security firm WhiteScope report having discovered over 8,600 security problems in some seven pacemakers under investigation. If malicious parties take advantage of these, the consequences could be deadly.
The researchers report having scrutinised a total of seven systems from four different manufacturers. They include pacemakers, implantable cardioverter-defibrillators, pulse generators and devices for treating heart rhythm disturbances. In all cases, the researchers talk about pacemakers and note that most of these devices contain the same basic components: a pacemaker, a patient monitoring device, a cloud-based infrastructure to transmit data to the doctor and a programming device for programming the pacemaker.
Purchasing programming equipment
Programming devices for pacemakers from all four manufacturers surveyed are 'controlled distribution'. This means that the devices are officially only available through the manufacturer and must be returned to this company by users after use. In practice, however, the programming devices also appear to be for sale on online auction websites. The researchers claim that the equipment is available for $500 to $3,000. Monitoring equipment for pacemaker can be bought for an amount ranging from $15 to $300, while pacemakers themselves can be purchased through secure sites for $200 to $3,000.
The researchers report having discovered a total of 8,600 known security problems in the seven pacemakers examined. This is possible because software on the systems is not kept up-to-date and is often outdated. These updates are used by software manufacturers to fix security problems, among other things. Those who do not install these updates leave these security holes wide open and give cybercriminals free rein. In the case of a pacemaker, this can have deadly consequences.
Unencrypted patient data
WhiteScope additionally reports having found patient data on a programming device purchased through online auction sites in two cases. These included Social Security numbers (the US counterpart of the Dutch Burger Service Nummer), names, phone numbers and medical data. This data was stored unencrypted on the programming equipment and was therefore accessible to anyone who managed to get their hands on it. The researchers therefore argue that hospitals need to be more careful with medical equipment they no longer use. The data came from a large hospital located on the US east coast.
Also noteworthy is that no programming equipment requires logging in to use these devices. This is not without risk, as it allows a thief to use a stolen device to programme pacemakers. In addition, programming devices attached to a pacemaker are not authenticated. This means that attackers who get hold of such devices and manage to get close enough to a patient with a pacemaker can modify the pacemaker's operation. Also, with programming equipment from a particular manufacturer, it is possible to program any pacemaker from that manufacturer. Pacemakers can only be programmed if they are in close proximity to the programming equipment.
Names of manufacturers not disclosed
The researchers did not disclose the exact devices and manufacturers involved. This is a deliberate choice to avoid putting patients with a pacemaker from these manufacturers at risk by publishing the research results. However, the researchers did contact the manufacturers and reported their findings so that the manufacturers can fix the problems. In addition, the problems were reported to the Industrial Control Systems Cyber Emergency Response Team (ICS-CERT), a US government agency that analyses security problems and alerts parties to them.
Given the similarities between the various systems studied, WhiteScope hopes that pacemaker manufacturers will work together to increase the digital security of their devices. The researchers argue that these manufacturers should compete on user experience and health aspects, not cybersecurity.
By: Wouter Hoeffnagel
Source: WhiteScope